Notes for DEWETRON-TRION-Applications-8.3.1.11034.exe
## 2026-11-13 Release Notice DEWETRON TRION API R8.3.1
Revision: 29aed4dbb8365b7c6e6427add05cb9f911ca9484
Components | Version
--------------------------------|--------------------------
DEWETRON Explorer | 8.3.1.11034
DEWETRON TRION API | 8.3.1.11034
DEWETRON TRION driver | 8.3.1.11034
DEWETRON DEWE2 driver | 8.3.1.11034
## Files
DEWETRON-TRION-Applications-8.3.1.11034.exe
## Requirements
* Update all TRION modules to the latest firmware included in this release
* Use OXYGEN R8.3.0 or higher
* Windows 10 64-Bit or better
## Installation procedure
* Install TRION applications
* Reboot system (if requested by the installer)
* Start DEWETRON Explorer and update each TRION board firmware to latest version (right-click on enclosure name)
* Power-cycle TRION boards (Power down system, then restart)
* Start DEWETRON Explorer and check for a successful firmware upgrade
## Fixes and Changes
* TRION3-AOUT: Fixed an issue where the analog output kept outputting the signal after the channel was switched off
* TRION3-CAN-FD: Fixed an issue where the CAN data stream stopped or did not start on some cards
* TRION Multi: Fixed voltage excitation not being switched off on disabled channels
* TRION-2402-MULTI and TRION-2400-dSTG: Fixed an issue where the bridge resistance could not be set after selecting a quarter bridge
* CHASSIS-CONTROLLER: Fixed clock drift on slave segments when the sync cable is removed and inserted
* Explorer: Added support for DEWE3-M12S
## New Firmwares
HW Support TRION3 modules | Firmware
--------------------------------|-----------
TRION3-CAN-FD-4 | FW: 0032
HW Support Chassis modules | Firmware
--------------------------------|-----------
TRION3-CONTROLLER | FW: 0023 |