Release Notice ORION DAQ Driver v2.3.1.0
2018-07-13
* Version: 2.3.1.0
* Components:
** 32-bit DAQ driver
** 64-bit DAQ driver
* New Features:
** supports Win10 64-bit systems (Winx64)
** supports Win7 32-/64-Bit systems
* Bugfixes:
** FW version number indication fixed for Orion 0816-1000N and Orion 1616-1000 (correct "101", wrong "901")
** ORION co-installer fixed to support Win10x64
** external clocking with DEWE-Clock fixed
* Known Limitation:
** At all times only one instance of the DAQ driver is allowed.
** Installing the driver without Administrator privileges will result in an incorrect installation
** When Orion 1616-50x, Orion 0816-100x or Orion 1616-100x are used in the system, the DMA buffer size has to be adjusted manually in order to support all channels of the cards. For the detailed descritpion please refer to <<Application_Note_64 Bit Driver.pdf>>
** Measurement start: If no CAN frames are received before the internal 24-bit counter overflows (~3 minutes @100 Hz), the overflow is not considered and the CAN timestamps are wrong. Workaround: add a CAN channel that always sends data. (problem @ BOSCH, will be fixed for ORION and Oxygen, was already fixed for TRION, won´t be fixed for ORION and DS => DS issue) |